SHIPS VIA FREIGHT (charges apply)
The Yamato SI series consists of economical dry heat sterilizers built with natural convection airflow. The SI dry sterilizer models can be used for samples that are sensitive to moisture such as powders, fat, & oils. Thanks to its non-corrosive feature, the SI dry-heat sterilizer can also be used for metal & instruments that may become corroded. The Yamato SI series static-air dry sterilizer guarantees massive savings with a much lower initial cost, maintenance cost, and operation cost.
Yamato’s programmable SI dry sterilizers deliver better operability with an improved temperature controller. For ultimate convenience, common operation settings including quick auto-stop & auto start operations can be easily accessed in the digital control panel. In addition, the SI series are programmable up to 30 steps, 6 patterns, and repeat operation so you can save effort and time. The SI dry sterilizer series are built with self diagnostic functions for ensured safety. Perform cost-effective dry-heat sterilization with Yamato’s SI401 77L dry sterilizer.
Operation and functions
- Improved temperature controller with better operability
- Digital setting by main menu and ▲▼ keys for fixed temperature, program, quick auto stop, auto stop and auto start operations
- Standard equipped with calibration off-set, lock function, power failure recovery mode
- Programmable with maximum 30 steps, 6 patterns, repeat operation
- Increased Safety Features
- Standard equipped with self diagnostic functions, earth leakage breaker and overheat prevention device
Specifications
Circulation Method | Natural convection dry laboratory sterilizers |
Operating Temp. Range* | Room temp. +5~260°C |
Temp. Adjustment Accuracy | 2°C (at 260°C) |
Temp. Distribution Accuracy | 20°C (at 260°C) |
Max. Temp. Reaching Time | ~70min. |
Interior / Exterior | Stainless steel / Chrome free electrogalvanized carbon steel sheet coated with chemical-proof baked-on finish |
Heater | Iron chromium wire heater / 1.36kW |
Sensor | K type Thermocouple |
Exhaust Port | I.D.30 mm x 2 (top) |
Temp. Controller | PID control |
Temp. Display | Measured temp. display: green 4 digit LED digital display / Setting temp. display: red 4 digit LED digital display |
Timer | 1 min~99hr 59 min and 100~999 hrs 30 min. (with timer wait function) |
Operation Function | Fixed temperature, quick auto stop, auto stop, auto start, program (6 pattern: 30 steps x 1, 15 steps x 2, 10 steps x 3, pattern repeat function) |
Additional Functions | Lock function, auto recovery after power failure, calibration off-set |
Safety Device | Self diagnostic functions (temp. sensor failure, heater disconnection, SSR short circuit, memory error, internal communication error, measurement temperature error, automatic overheat prevention), overheat prevention device, earth leakage breaker |
Internal Dimensions WxDxH | 600 x 530 x 500 mm |
External Dimensions WxDxH | 700 x 640 x 877 mm |
Internal Capacity | 159L |
Shelf Plate With Standard Load | ~15 kg/pc. |
Shelf Rest Step Number / Pitch | 13 steps / 30 mm |
Power Source (50/60Hz) | AC115V 12.5A (with plug) |
Weight | ~59 kg |
Shelf Plate Material | Stainless punching metal |
Shelf Plate / Bracket | 2pcs./4pcs. (bottom one is screwed) |