Features
● Comes with 48-well design and two models.
● 4-channel 48E can detect up to 4 target types per tube at the same time, while 2-channel 48R can detect up to 2 targets.
● Efficient temperature control with only taking40 minutes to complete a standard PCR amplification program.
● Temperature accuracy is controlled within 0.1℃.
● In 2 seconds, all 48-hole 4(48E)-/2(48R)-color fluorescence scanning can be completed.
● Designed with a built-in 7-inch full-color touch screen, independent control system and USB port for data export with the storage of at least 1000 experiments.
● For the stand-alone operation the device can also be connected to an optional thermal printer for direct printing of sample amplification curves and CT values.
● Compatible with the fluorescence analysis functions of a variety of isothermal amplification technologies.
● Independent data management software built in Gentier 48E/48R for the centralized management and analysis of your massive experimental data, to minimize the errors caused by manual analysis of original data.
Model |
48E |
48R |
Throughput |
48 |
48 |
Fluorescence Channels |
4 |
2 |
Compatible Fluorophores |
Channel 1: FAM, SYBR Green I, etc. Channel 2: VIC, HEX, TET, JOE, etc. Channel 3: ROX, Texas Red, etc. Channel 4: Cy5, etc. |
Channel 1: FAM, SYBR Green I, etc. Channel 2: VIC, HEX, TET, JOE, etc. |
Light Source |
High-brightness, long-life and maintenance-free LED light source |
|
Detector |
Photodiodes (PDs) |
|
Fluorescence Scanning Time |
48-well 4-channel detection completed within 2s |
48-well 2-channel detection completed within 2s |
Heating Rate |
Maximum heating ramp rate ≥8.0℃/s; Average heating ramp rate ≤6.2℃/s |
|
Cooling Rate |
Maximum cooling ramp rate ≥6.2℃/s; Average heating ramp rate≥4.5℃/s |
|
Temperature Accuracy |
≤ 0.1℃ |
|
Special Temperature Setting Function |
Support thermal gradients PCR, long PCR, Touch Down PCR |
|
Suitable Consumables |
0.2ml clear non-skirt 48-well plates, 8-tube strips, single tubes, etc. |
|
Sample Testing linearity and Repeatability |
linear Correlation: CV≥0.999 Repeatability: cycle threshold (Ct) value CV ≤0.5% |
|
Software Analysis |
Qualitative analysis, absolute quantitative analysis, relative quantitative analysis, end point fluorescence analysis, melting curve analysis, and genotyping analysis, etc. |
|
Control Method |
Stand-alone operation: 7-inch touchscreen control; Cloud-enabled: PC software control via direct connection or LAN (local area network); Up to 10 devices controlled by one PC |
|
Power Failure Protection |
Automatic recovery of the experiment and other functions when the power is on again after cutting off, without waiting for the power-on of the computer or software control |
|
Data Storage and Transmission |
A single machine can store more than 1,000 experimental data files, which can be imported and exported via USB disks |
|
Stand-Alone Analysis Results |
No need for an extra computer: experimental data analysis enabled within the instrument; Direct printing function (optional) to output the sample amplification curve and CT value, simply by connecting to the compatible thermal printer |
|
Reporting Function |
Built-in experiment report templates for a variety of industries; Fully open universal reporting of which the contents and formats can be customized |
|
Operating System |
Windows XP Pro or higher |
|
Power Supply and Power Consumption |
AC 100-240V, 50-60Hz; 600VA |
|
Weight |
11Kg (net) |
|
Instrument Dimension |
260x400x260mm |